SLICE-DLC - Dual Laser Controller

Dual laser current and temperature controller family. 200 or 500 mA maximum per current channel, configuration dependent

SKU: VES-SLICE-DLC Category:
Contact us to inquire about this product

If you love this product and wish for a customized quote contact us at number +34 91 3854 395 and we will be happy to provide you with more info!

Product overview

SLICE-DLC is a Vescent dual laser current and temperature controller family. This exact Excel reference identifies the documented product or family. Current, voltage, bandwidth, optical power and wavelength are specified with their configuration and test conditions. For families, select the exact option with IberOptics; compatible accessories are identified by their own function.

Documentation

Vescent SLICE-DLC family technical documentation

ManufacturerVescent
Family documentationView family documentation
Product typeDual laser current and temperature controller family
Control channels2 current channels; 4 temperature channels
Maximum output current200 or 500 mA maximum per current channel, configuration dependent
Model configuration details
Product type Dual laser current and temperature controller family
Current Channels per Unit DLC-200 Value: 2; DLC-500 Value: 2
Current Capacity per Channel DLC-200 Value: 200; DLC-500 Value: 500; Units: mA
Minimum Deliverable Current DLC-200 Value: 0; DLC-500 Value: 0; Units: mA
Polarity DLC-200 Value: Cathode ground only; DLC-500 Value: Cathode ground only
Current Noise Density DLC-200 Value: ≤100; DLC-500 Value: ≤200; Units: pA/√Hz; Comments: At > 1 kHz
Current Noise DLC-200 Value: ≤50 ≤100 ≤300; DLC-500 Value: ≤100 ≤150 ≤500; Units: nA, rms; Comments: Integrated from 10 Hz to 100 kHz Integrated from 10 Hz to 1 MHz Integrated from 10 Hz to 10 MHz
Current Setpoint Resolution DLC-200 Value: ≤0.0002; DLC-500 Value: ≤0.0005; Units: mA
Temperature Drift DLC-200 Value: ≤1; DLC-500 Value: ≤1; Units: µA/°C
Compliance Voltage Value: ≤18; Units: V; Comments: automatically adjusted
Servo Input Bandwidth DLC-200 Value: ≥10; DLC-500 Value: ≥10; Units: MHz
RF Input Modulation Bandwidth DLC-200 Value: 1; DLC-500 Value: 1; Units: GHz
LIV Acquisition Range DLC-200 Value: 0 to Imax; DLC-500 Value: 0 to Imax; Units: mA
LIV Acquisition Time DLC-200 Value: 1 to 100; DLC-500 Value: 1 to 100; Units: s
Total Temp. Channels Value: 4; Comments: 2 per current channel
Compatible Transducer Value: TEC or resistive heater
Compatible Sensor Value: NTC thermistor
Temperature Stability Value: ±0.2; Units: mK; Comments: w/ properly designed temperature plant
Precision Value: ±0.2; Units: mK
Control Capacity Value: 20; 40; Units: W; Comments: per channel; total
Current Capacity Value: 5; Units: A; Comments: per temperature plant
Input Impedance Value: 50, 1M; Units: Ω; Comments: Configurable
Input Voltage Limit Value: ±10; Units: V
Output Voltage Range Value: ±10; Units: V
Overvoltage Category Value: 2
Input Line Voltage Value: 100 - 240; Units: VAC
Mains Supply Voltage Fluctuations Value: ±10; Units: %
Frequency Value: 50 - 60; Units: Hz
Phase Value: 1 phase
Maximum Input Current Value: 2; Units: A
User-Serviceable Fuse Value: HBC 2.0 A 250 V
Indoor or Outdoor Use Min. Value: Indoor Use Only; Comments: Dry Location Only
Pollution Degree of Environment Max. Value: 2
Operating Altitude Max. Value: 10000 (3048); Units: ft (m)
Operating Temperature Min. Value: 15; Max. Value: 30; Units: °C
Operating Relative Humidity Min. Value: 20; Max. Value: 60; Units: %; Comments: Non-condensing
Dew Point Max. Value: 15; Units: °C
Controller channels 2 current channels; 4 temperature channels
Output current 200 or 500 mA maximum per current channel, configuration dependent

I have read and accepted your terms and conditions and privacy policy