GEMINI-2D Interferometer

Phase-stable interferometer for two-dimensional electronic spectroscopy; 400–2300 nm standard; 250–3500 nm ultra-broadband configuration

SKU: NRS-GEMINI-2D Interferometer Category:
Contact us to inquire about this product

If you love this product and wish for a customized quote contact us at number +34 91 3854 395 and we will be happy to provide you with more info!

Product overview

GEMINI-2D is the NIREOS interferometer for time-resolved and two-dimensional electronic spectroscopy. It generates two collinear, phase-locked pump pulses while keeping dispersion and one absolute arrival time constant during the delay scan. S/L and symmetric/asymmetric configurations have different scan ranges; this Excel family SKU does not create separate undocumented ordering references.

Documentation

NIREOS GEMINI-2D manufacturer family brochure

ManufacturerNIREOS
Spectral range and conditions400–2300 nm standard; 250–3500 nm ultra-broadband configuration
Family documentationView family documentation
Dimensions180 × 180 × 90 mm
Product typePhase-stable interferometer for two-dimensional electronic spectroscopy
Weight and configuration2 kg
Model configuration details
Function Phase-stable interferometer for two-dimensional electronic spectroscopy
Spectral range and configuration 400–2300 nm standard; 250–3500 nm ultra-broadband configuration
Dimensions 180 × 180 × 90 mm
Weight 2 kg
Product scope Optical interferometer module. It does not include an image sensor or independently produce a camera pixel resolution/frame rate. The external detector/camera and acquisition configuration determine those values.
Pulse-pair generation Two collinear, phase-locked replicas of the input ultrashort pulse; constant introduced dispersion and fixed absolute arrival time of one replica during the delay scan.
Relative delay at 600 nm, symmetric configuration S: −400 to +400 fs; L: −1050 to +1050 fs
Relative delay at 600 nm, asymmetric configuration S: −100 to +700 fs; L: −100 to +2000 fs
Relative delay stability <1 attosecond; distinct from spectral resolution or acquisition rate.
Scan mode Step scan; delay dwell time selected through software.
Spectral resolution Depends on selected delay range and wavelength; consult the manufacturer resolution graph for the selected S/L and symmetric/asymmetric configuration.

I have read and accepted your terms and conditions and privacy policy