BM-A-15W-14-TX-ENH

1 mW–15 W, continuous average power; 0.25–40 µJ at 1064 nm, pulse duration <10 ns; 20–1000 µJ at 1064 nm, pulse duration 10 ms

SKU: LPT-BM-A-15W-14-TX-ENH Category:
Contact us to inquire about this product

If you love this product and wish for a customized quote contact us at number +34 91 3854 395 and we will be happy to provide you with more info!

Product overview

BM-A-15W-14-TX-ENH is a high-sensitivity BLINK thermal laser measurement reference. Pulse-energy limits depend on wavelength and pulse duration; the continuous power limit is separate from the brief overload rating. This ENH reference includes the HSM-1000 enhanced high-speed meter.

Documentation

LaserPoint high-speed thermal sensors family catalogue — 2024 revision 1

ManufacturerLaserPoint
CoolingForced air; recommended ambient 10–35 °C
Spectral range and conditions0.2–11 µm absorber range; calibration wavelengths selected separately
Data interfaceHirose IX sensor connection to HSM-1000
Family documentationView family documentation
Dimensions60 x 60 x 66 mm
Product typeHigh-speed thermal power/energy sensor and HSM-1000 enhanced meter system
Measured power and conditions1 mW–15 W, continuous average power
Measured energy and conditions0.25–40 µJ at 1064 nm, pulse duration <10 ns; 20–1000 µJ at 1064 nm, pulse duration 10 ms
Active aperture14 × 14 mm; minimum beam diameter 3 mm
Measurement uncertainty±5% power / ±5% energy calibration uncertainty
Repetition rate and conditionsUp to 1 MHz, within documented pulse-energy and average-power limits
Model configuration details
Max. Average Power 15 W
Power Calibration Uncertainty ± 5%
Max repetition rate 1 Mhz
Energy Resolution 0.01 µJ
Calibration Uncertainty ± 5%
Aperture 14 x 14 mm
Type T
Spatial Uniformity (1) ± 5 %
Min. Beam diameter 3 mm
Absorber Spectral Range 0.2 - 11 µm
Calibration Spectral Range 0.355 µm 0.532 µm 1.07 µm 10.6 µm (c) (d)
Max Power Density (2) 0.1 kW/cm²
Max Energy Density (3) 35 mJ/cm²
Cooling Forced Air (e)
Maximum sensor operating temperature 40 °C
Dimensions 60 x 60 x 66 mm
Cable length 2 m
Stand and Post Light Duty Stand
(1) 3mm beam diameter, scanning 80% of active area (2) Damage thresholds also depend on power level. (3) Single shot (a). wavelength 1064nm, pulse duration < 10 ns (b). wavelength 1064nm, pulse duration = 10 ms (c). @ 10.6 µm sensor reflectivity 70% (d). Others wavelength on request (e). Recommended ambient temperature: 10 -35 °C
Pulse energy and measurement conditions 0.25–40 µJ at 1064 nm, pulse duration <10 ns; 20–1000 µJ at 1064 nm, pulse duration 10 ms
Calibration and absorber Absorber range is not the calibration range. Standard listed calibrations: 355, 532 and 1070 nm; 10.6 µm option has 70% reflectivity; other wavelengths on request.
Cooling and conditions Forced air; recommended ambient 10–35 °C
Meter configuration HSM-1000 included in the ENH system; standard sensor needs compatible meter
Maximum transient average power 20 W for at most 2 s and at most 20% duty cycle