A-1200-D60-SHC

Thermal Sensors; model-specific configuration and conditions are listed in the technical table.

SKU: LPT-A-1200-D60-SHC Category:
Contact us to inquire about this product

If you love this product and wish for a customized quote contact us at number +34 91 3854 395 and we will be happy to provide you with more info!

Product overview

A-1200-D60-SHC is a thermal sensors from LaserPoint. The technical configuration table below is taken from the manufacturer's page for this model. Values, alternative modes, wavelength-specific performance, measurement conditions and option codes must be read together. The current model-specific manufacturer page is the source for the listed specifications; the shared 2023 catalog is historical family documentation and covers this ordering code. Where revisions differ, use the current model table and confirm the selected configuration with IberOptics. The image is the manufacturer’s product or family photograph/diagram linked from this model page and does not imply that optional accessories are included. Availability must be confirmed with IberOptics.

Documentation

LaserPoint product-family catalog (2023, Rev. 1; historical manufacturer document)

ManufacturerLaserPoint
CoolingCooling: Forced Air with Fan (a)
Spectral range and conditionsAbsorber Spectral Range: 0.19 - 11 µm
Data interfaceCable length - connector: 5.0 m - DB15 5.0 m – USB (U option) 5.0 m - RS232 (R option)
Family documentationView family documentation
Product typeThermal Sensors
Manufacturer document and hardware revisionLaserPoint product-family catalog (2023, Rev. 1; historical manufacturer document)
Measured power and conditionsMax. Average Power: 1200 W
Measured energy and conditionsMax. Energy (with 100 ms pulse): 1200 J
Response time and conditionsResponse Time (0-90%): 6 sec
Active apertureAperture: 60 mm
Measurement uncertaintyPower Calibration Uncertainty: ± 3%
Model configuration details
Power Mode
Max. Average Power 1200 W
Max. Average Power (1) 1200 W
Min. Power 2 W
Power Resolution 100 mW
Noise Equivalent Power (NEP) 100 mW
Response Time (0-90%) 6 sec
Power Calibration Uncertainty ± 3%
Power Linearity (2) ± 1.5%
Single Shot Energy Mode
Max. Energy (with 100 ms pulse) 1200 J
Min. Energy 5 J
Energy Resolution 100 mJ
Energy Calibration Uncertainty ± 5%
Absorber Specs
Aperture 60 mm
Type SHC
Absorber Spectral Range 0.19 - 11 µm
Calibration Spectral Range 0.2 - 1.1 µm, 10.6 µm
Max Power Density (3) 5 kW/cm² @1 kW
Max Energy Density (3) 5ms pulse width: 115 J/cm²
10µs pulse width: 4 J/cm²
10ns pulse width: 1 J/cm²
General Characteristics
Cooling Forced Air with Fan (a)
Weight 4.4 kg
Dimension 143 x 143 x 132 mm
Cable length - connector 5.0 m - DB15
5.0 m – USB (U option)
5.0 m - RS232 (R option)
Stand and Post Heavy Duty Stand
Notes
(1). 2 minutes max
(2). Detector centrally irradiated @50% of useful surface.
(3). Damage thresholds also depend on power level. Please see damage graphs for more details.
(a). 24V DC Power Supply Included

I have read and accepted your terms and conditions and privacy policy